Medical sealing machine malfunction and solution.

Date of release:2018-02-05 Author: Click:


Medical sealing machine.jpg

The fault phenomenon The cause of the problemThe solution
 Unable to boot normally 

1. Check whether the 220V ac is connected.

2. Check whether the power switch is in place.

3. Whether the fuse in the switch is damaged.

4. Check whether the heating plate is damaged by 

self-recovery insurance.

5. Check whether the dc power supply is damaged.

1. Connect to normal 220V ac.

2. Repeatedly "switch off" power switch.

3. Replace the 5A20mm fuse with a screwdriver with the fuse holder. (the device manual has a spare fuse in the bag.)

4. After the equipment is cut off, use a multimeter resistance file to check whether the circuit breaker is broken. If it is normal, contact the manufacturer for maintenance.

5. Check whether there is current output or replace dc power supply.

 The print head does not print.

1. Check whether the printing content is set correctly.

2. Check whether the ribbon is depleted or damaged.

3. Check whether the corresponding wiring is connected correctly.

4. Check whether the color belt motor turns normally.

5. Check whether the photoelectric switch has signal input.

6. Check whether the print head is burned.

7. Check whether the motherboard element is damaged.

 

1. Reset the print content.

2. Replace the new ribbon.

3. Connect the ribbon motor and print head and main board correctly.

4. If the ribbon motor is damaged, replace the ribbon motor.

5. Check whether the printed signal has output or replace the photoelectric switch.

6. Replace the print head.

7. Replace the motherboard.

 Typing is not complete.

1. Check whether the print head is missing the needle.

2. Check whether the corresponding needle attachment of the circuit board is loose.

3. Check whether the print head is burned.

4. Check whether the main board components are damaged.

1. Check whether the printing head is missing the needle, if the needle is missing, replace the print head.

2. Connect the connecting line between the print head and the main board.

3. Replace the print head.

4. Replace the motherboard.

 Typing is not clear.

1. Check whether the ribbon is consumed or damaged.

2. Check whether the screw of the printing head is loose.

3. Check whether the wiring of the ribbon motor is loose.

4. Check whether the ribbon motor is damaged.

 

1. Replace the new ribbon.

2. Tighten the printing head to loosen screws.

3. Connect the connection between the ribbon motor and the motherboard to make it reliable.

4. Replace the ribbon motor.

 

Print head overheating

 The control panel displays the print head overheating tips. The device is switched on after cooling.
 Turn on the motor.

1. Check whether the wiring connection is reliable.

2. Check whether there is voltage input for 220V.

3. Check whether the synchronous motor is damaged.

4. Check whether there is any problem with the circuit board.

 

1. Reconnect the connection between the synchronous motor and the motherboard.

2. Access to normal 220V ac.

3. Replace the synchronous motor.

4. Replace the motherboard.

 The motor does not turn when the temperature is set 

1. Check whether the entrance photoelectric switch has a signal.

2. Check whether the motherboard is damaged.

1. Adjust the position of the photoelectric switch or replace the photoelectric switch.

2. Replace the motherboard.

 

The fault phenomenon
 The cause of the problem The solution
 

The heating plate is not heated.

1. Check whether the temperature setting is correct.

2. Check whether the wiring is reliable.

3. Check whether there is voltage input for 220v.

4. Check whether the heating board is damaged or not.

5. Check whether the heating panel is damaged.

6. Check the temperature sensor for damage.

7. Check whether the main board components are damaged.

 

1. Reset the heating temperature.

2. Reconnect the connection between the heating plate and the motherboard.

3. Access to normal 220V ac.

4. Replacement of insurance

5. Replace the heating plate.

6. Replace the temperature sensor.

7. Replace the motherboard.

 Heating plate temperature cannot be raised to set temperature.

1. Check whether the temperature sensor is damaged.

2. Check whether the main board is damaged.

 

1. Replace the temperature sensor resistance.

2. Replace the motherboard.

 The heating plate is always heated and cannot be controlled. 

1. Check whether the wiring is reliable.

2. Check whether the temperature sensor is damaged.

3. Check whether the main board is damaged.

 

1. Reconnect the connection between the heating plate and the motherboard.

2. Replace the temperature sensor resistance.

3. Replace the motherboard.

 A sensor failure occurred. 

1. Check whether the wiring is reliable.

2. Whether the resistance of temperature sensor is normal.

3. Check whether the main board is damaged.

 

1. Reconnect the connection between the temperature sensor and the motherboard.

2. Replace the temperature sensor resistance.

3. Replace the motherboard.

 The display screen appears with a splash screen or messy code.

1. Program execution chaos.

2. Check whether the control screen is damaged.

3. Check whether the main board is damaged.

1. Restart the power switch.

2. Replace the control panel.

3. Replace the motherboard.

 The display screen is not displayed after power on.

1. Check whether the ac power supply is normal.

2. Check whether the dc power supply is normal.

3. Check whether the main board is damaged.

4. Check whether the control screen is damaged.

1. Access to normal 220V ac.

2. Replace dc power supply.

3. Replace the motherboard.

4. Replace the control panel.

 The sealing is not good enough.

1. Appearance quality problem of sealing surface.

2. The plastic film creases and melts.

3. The seal is not fastened properly.

4. The thickness of plastic film is large.

5. The pressure in the device is reduced.

6. There is water and stains on the seal.

7. The pressure roller is stained with dirt.

There is something foreign on the seal.

 

1. Set the sealing temperature to increase or decrease.

2. Lower the temperature setting value of the seal.

3. Raise the temperature setting value of the seal.

4. Raise the temperature setting value of the seal.

5. Check the pressure belt wheel and adjust the compression spring properly.

6. Remove moisture and stains from the seal.

7. Remove the pressure wheel fillet.